| Sensors and sample handling system |
|
| Number of sensors |
4, also possible using only 1, 2 or 3 sensors |
| Flow profile |
Parallel plate flow |
| Typical sample volume |
500 µl per sensor |
| Volume above sensor |
40 µl |
| Working temperature |
18-50 ° C |
| Temperature stability |
± 0.02 ° C |
| Flow rates |
0-800 µl/min |
| Cleaning |
All parts easily removed for separate cleaning |
| |
|
| Frequency and dissipation characteristics |
|
| Frequency range |
1-70 MHz |
| Maximum measurement rate: |
Up to 200 data points / s |
| Typical mass sensitivity in liquid*: |
1.8 ng / cm2 (0.1 Hz) |
| Typical detectable change in D in liquid * |
0.1 x 10-6 |
| |
|
| Software capabilities |
|
| General |
Plots, data editing and calculations |
| Ouput data, Sauerbrey |
Mass & thickness vs time |
| Output data, Voight viscoelastic model |
Thickness, viscosity & elasticity vs time |
| Other functions |
Kinetic constants, temperature ramps, own models |
| |
|
| Dimensions |
|
| Electronics unit |
18 x 36 x 21 cm (HxWxD), 9 kg |
| Measurement chamber |
12 x 23 x 34 cm (HxWxD), 8 kg |